Analysis of albumin Raman scattering registration efficiency from different volume and shape cuvette

Lykina, Anastasia and Artemyev, Dmitry and Bratchenko, Ivan (2017) Analysis of albumin Raman scattering registration efficiency from different volume and shape cuvette. Journal of Biomedical Photonics & Engineering, 3 (2). 020309. ISSN 24112844

[thumbnail of 3157-8818-1-PB.pdf] Text
3157-8818-1-PB.pdf - Published Version

Download (1MB)

Abstract

The cuvette shape influence on the Raman registration efficiency at small volumes of albumin (0.15-1 ml) using the PLS model was analysed. This paper examines the smallest measurement error RMSE of 3-6 g/l with a correlation coefficient R2 = 0.95 was achieved for pure Raman spectra extraction by baseline correction with asymmetric least squares smoothing method for aluminum spherical type cuvette-reflectors. Its curvature center was matched with the focal spot of the exciting laser. The error of albumin concentration determination equaled about 5% of albumin normal concentration in a blood plasma using a spherical cuvette with a volume of 0.55 ml. A further decrease of the cuvette volume led to a sharp loss of the method prediction accuracy.

Item Type: Article
Subjects: Research Scholar Guardian > Multidisciplinary
Depositing User: Unnamed user with email support@scholarguardian.com
Date Deposited: 11 Mar 2023 10:53
Last Modified: 19 Sep 2023 06:27
URI: http://science.sdpublishers.org/id/eprint/237

Actions (login required)

View Item
View Item